Clean equipment is vital to the pharmaceutical industry due to potential cross contamination between product runs of various drugs.

Total Organic Carbon (TOC) analysis is a widely accepted way to analyze for residual pharmaceutical components.

WOT-100 TOC analyzer adopts the UV light-activated persulfate oxidation method. 

TOC content is determined by first removing inorganic carbon (IC) from the sample through the addition of

phosphoric acid. The organic carbon is oxidized to carbon dioxide (CO2) using UV oxidation with

sodium persulfate. The CO2 produced is then swept to the non-dispersive infrared detector (NDIR) and

measured. 

Based on above steps, we can get the TC and TIC, thus the TC - TIC= TOC.

Features

UV light-activated persulfate oxidation method. 

Equipped with eight-way rotary valve for automatic acid intaking, oxidant intaking, sample intaking ,purge and waste discharge.

Poweful software compliance with  the requirements of 21CFR Part 11, audit trail, CSV(Computerized System Validation).

Remove halide and bicarbonate interference;

Low cost, main consumables are Phosphoric acid solution and Sodium persulfate solution;

Wide detemination range.

Easy to calibration, support the one point and multi-point calibration.

Built-in aeration tank, when testing NPOC, a series of processes such as aeration, acid addition and sample injection are completed by the instrument automatically.


Technical Parameters

Power supply:          (220±10%)VAC 50/60Hz

Power consumption: 300W
Indication error: ±5%
Repeatability: RSD≤3%
Determination Range:(0-100.000)mg/L
Analysis time: 4 min.
Required sample temperature:(1-95)℃
Ambient temperature:(10-40)℃

Noise:≤55dB(A)

Dimensions: 55cm×50cm×53cm

Net Weight: 35Kgs

Applications

Cleaning validation in pharmaceutical industry

 

 